Development of solid-state silicon devices as ultra cold neutron detectors
journal contribution
posted on 2023-06-07, 19:15 authored by C A Baker, K Green, M G D van der Grinten, P S Iaydjiev, S N Ivanov, S Al-Ayoubi, P G Harris, J M Pendlebury, D B Shiers, P GeltenbortWe present a study aimed at developing and optimising methods for ultra-cold neutron (UCN) detection using solid-state silicon detectors. For this we investigated the characteristics of neutron to charged particle converters ((6)Li, (10)B) and compared different deposition techniques. We find that a deposition of 600 µg/cm(2) of (6)LiF is required to ensure maximum UCN detection efficiency using (6)Li as converter. The UCN detectors that we have developed are now fully operational as a flux monitor and as a UCN spectrum analyser in the neutron electric dipole moment experiment at the Institut Laue Langevin in Grenoble.
History
Publication status
- Published
Journal
Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated EquipmentISSN
0168-9002Publisher
ElsevierExternal DOI
Issue
3Volume
487Page range
511-520Pages
1.0Department affiliated with
- Physics and Astronomy Publications
Full text available
- No
Peer reviewed?
- Yes