posted on 2023-06-07, 07:59 authored by Chantelle RizanChantelle Rizan, Mahmood BhuttaMahmood Bhutta
Background Hybrid surgical instruments contain both single-use and reusable components, potentially bringing together advantages from both approaches. The environmental and financial costs of such instruments have not previously been evaluated. Methods We used Life Cycle Assessment to evaluate the environmental impact of hybrid laparoscopic clip appliers, scissors, and ports used for a laparoscopic cholecystectomy, comparing these with single-use equivalents. We modelled this using SimaPro and ReCiPe midpoint and endpoint methods to determine 18 midpoint environmental impacts including the carbon footprint, and three aggregated endpoint impacts. We also conducted life cycle cost analysis of products, taking into account unit cost, decontamination, and disposal costs. Results The environmental impact of using hybrid instruments for a laparoscopic cholecystectomy was lower than single-use equivalents across 17 midpoint environmental impacts, with mean average reductions of 60%. The carbon footprint of using hybrid versions of all three instruments was around one-quarter of single-use equivalents (1756 g vs 7194 g CO2e per operation) and saved an estimated 1.13 e-5 DALYs (disability adjusted life years, 74% reduction), 2.37 e-8 species.year (loss of local species per year, 76% reduction), and US $ 0.6 in impact on resource depletion (78% reduction). Scenario modelling indicated that environmental performance of hybrid instruments was better even if there was low number of reuses of instruments, decontamination with separate packaging of certain instruments, decontamination using fossil-fuel-rich energy sources, or changing carbon intensity of instrument transportation. Total financial cost of using a combination of hybrid laparoscopic instruments was less than half that of single-use equivalents (GBP £131 vs £282). Conclusion Adoption of hybrid laparoscopic instruments could play an important role in meeting carbon reduction targets for surgery and also save money.

History

Publication status

  • Published

File Version

  • Published version

Journal

Surgical Endoscopy

ISSN

0930-2794

Publisher

Springer Science and Business Media LLC

Volume

36

Page range

4067-4078

Event location

Germany

Department affiliated with

  • Clinical and Experimental Medicine Publications

Full text available

  • Yes

Peer reviewed?

  • Yes

Legacy Posted Date

2023-01-04

First Open Access (FOA) Date

2023-01-04

First Compliant Deposit (FCD) Date

2022-12-20

Usage metrics

    University of Sussex (Publications)

    Categories

    No categories selected

    Licence

    Exports

    RefWorks
    BibTeX
    Ref. manager
    Endnote
    DataCite
    NLM
    DC