University of Sussex
Browse

File(s) not publicly available

Intramedullary nails with two lag screws

journal contribution
posted on 2023-06-08, 15:57 authored by C J Brown, Chang WangChang Wang, A L Yettram, P Procter
Objective. To investigate the structural integrity of intramedullary nails with two lag screws, and to give guidance to orthopaedic surgeons in the choice of appropriate devices. Design. Alternative designs of the construct are considered, and the use of a slotted upper lag screw insertion hole is analysed. Background. Intramedullary fixation devices with a single lag screw have been known to fail at the lag screw insertion hole. Using two lag screws is considered. It has also been proposed to use a slot in the nail for the upper lag screw to prevent the upper lag screw from sticking. Methods. Bending and torsion load cases are analysed using finite element method. Consideration of both load conditions is essential. Results. The results present the overall stiffness of the assembly, the load sharing between lag screws, and the possibility for cut-out to occur. Conclusions. While the slot for the upper lag screw might be advantageous with regard to the stresses in the lag screws, it could be detrimental for cut-out occurring adjacent to the lag screws.

History

Publication status

  • Published

Journal

Clinical Biomechanics

ISSN

02680033

Publisher

Elsevier

Issue

5

Volume

19

Page range

519-525

Department affiliated with

  • Engineering and Design Publications

Full text available

  • No

Peer reviewed?

  • Yes

Legacy Posted Date

2013-10-03

Usage metrics

    University of Sussex (Publications)

    Categories

    No categories selected

    Exports

    RefWorks
    BibTeX
    Ref. manager
    Endnote
    DataCite
    NLM