posted on 2023-06-09, 06:50 authored by C J Forman, H Tomes, B Mbobo, R J Burman, M Jacobs, Thomas BadenThomas Baden, J V RaimondoThe ability to reliably and precisely deliver picolitre volumes is an important component of biological research. Here we describe a high-performance, low-cost, open hardware pressure ejection system (Openspritzer), which can be constructed from off the shelf components. The device is capable of delivering minute doses of reagents to a wide range of biological and chemical systems. In this work, we characterise the performance of the device and compare it to a popular commercial system using twophoton fluorescence microscopy. We found that Openspritzer provides the same level of control over delivered reagent dose as the commercial system. Next, we demonstrate the utility of Openspritzer in a series of standard neurobiological applications. First, we used Openspritzer to deliver precise amounts of reagents to hippocampal neurons to elicit time- and dose-precise responses on neuronal voltage. Second, we used Openspritzer to deliver infectious viral and bacterial agents to living tissue. This included viral transfection of hippocampal interneurons with channelrhodopsin for the optogenetic manipulation of hippocampal circuitry with light. We anticipate that due to its high performance and low cost Openspritzer will be of interest to a broad range of researchers working in the life and physical sciences.
History
Publication status
- Published
File Version
- Published version
Journal
Scientific ReportsISSN
2045-2322Publisher
Nature Publishing GroupExternal DOI
Issue
1Volume
7Department affiliated with
- Neuroscience Publications
Full text available
- Yes
Peer reviewed?
- Yes